The following specifications apply to both, UT320 and UT340 pulser receivers.
Pulser Specifications
Standard
Fast
Pulse Specification
Rise time
<2.0 ns
<1.0 ns
Fall time
<2.0 ns
<1.0 ns
Load Impedance
50 ohms
50 ohms
Minimum Load Impedance
capacitive to 10 Amps
capacitive to 5.2 Amps
Pulse Voltage Control
Minimum Output Voltage
100 V
100 V
Maximum Output Voltage
500 V
250 V
Maximum Voltage
Increment
2 V
2 V
Pulse Width Control
Pulse Width Minimum
5.0 ns
2.0 ns
Pulse Width Maximum
80 ns
80 ns
Pulse Width Resolution
0.20-0.50 ns
0.20-0.50 ns
Pulse Width Accuracy
10%
10%
PRF Control
PRF Minimum
200 Hz
200 Hz
PRF Maximum
20 kHz
20 kHz
Receiver Specifications
Gain Specification
Bandwidth (-3 dB)
1 MHz to 150 MHz
Input Impedance
50 ohms
Equivalent Input Noise
100 µV p-p max
Maximum Output Level
0 dBm
Output Headroom
6 dB
Gain Control
Voltage Gain Minimum
63 dB
Voltage Gain Maximum
Gain Increment
1 dB
Interfaces
Serial Communications
RS-232 (connection to host computer)
RS-485 (connection to additional pulser receivers)
User Input/Output
Four programmable digital inputs and outputs
Two sets of programmable relay contacts
Dimensions and weight
Height
86.6mm (3.5")
104mm (4") - including removable feet
Width
428.6mm (17") - desktop version
482mm (19") - rackmount version including side handles
Depth
507mm (20") - including front panel BNC connectors protruding 14mm and back panel cord wraps protruding 25mm
Weight
13.7kg (30.1lbs)
Product Ordering Options
The following options can be specified when ordering UT320 and UT340 instruments:
Rack mounting option
Consists of modified front panel design and an option to route front BNC connections to the back panel.
Fast Pulser option
Gives you pulse rise and fall times that are under 1ns.
Continuously Variable Filter option
Provides continuously variable second order low-pass and high-pass filtering as well as selectable RF signal inversion. Read more...